@article{oai:kumadai.repo.nii.ac.jp:00024763, author = {中村, 祐也 and Nakamura, Yuuya and 冨口, 静二 and Tomiguchi, Seiji and 勝田, 昇 and Katsuda, Noboru and 鷹本, 慶子 and Takamoto, Keiko}, issue = {7}, journal = {日本放射線技術学会雑誌}, month = {Aug}, note = {The aim of this study was to evaluate the effect of reconstruction parameters on the measurement of global left ventricular (LV) volume and define the appropriate reconstruction parameters when using the ordered subsets expectation maximization (OS-EM) method and 3D OS-EM (AstonishTM) method including the collimator distance response (CDR) recovery (RC) for myocardial perfusion SPECT study. Method: An anthropomorphic torso phantom with a 56 ml LV part was used. The LV volume was calculated with QPS software by the Update number (iteration number×subsets number) of OS-EM and 3D OS-EM (AstonishTM). Results and Conclusion: LV volumes calculated with OS-EM and AstonishTM without attenuation and scatter corrections corresponded to the true obtained by the Update number about 32 times and 24 times when using the OS-EM and AstonishTM method with attenuation and scatter corrections, respectively. However, LV volumes have changed greatly in the AstonishTM method according to the change in the Update number. Appropriate numbers of iterations and subsets are the measurement of global LV volumes, especially when using the OS-EM algorithm with RC.}, pages = {931--937}, title = {補正組み込み型OS-EM法の再構成条件が左室内腔容積算出に及ぼす影響}, volume = {65}, year = {2009}, yomi = {ナカムラ, ユウヤ and トミグチ, セイジ and カツダ, ノボル and タカモト, ケイコ} }